Can I adjust the brightness of the screen?

Only on the InBody230/320.

Depending on the year the unit was manufactured, there will be either a knob or a hole behind the head of the unit next to the USB port.  You can either turn the knob for brightness or you will need a Philips screw driver to turn the knob internally.

How do I upload a logo to the InBody unit?

Please email us at CS@inbodyusa.com with your logo and indicate which InBody model you own.  We will then format the logo and email it back to you. Once you receive your formatted logo, follow the instructions below on how to upload the logo

 

1. Drag the “logo” file onto your InBody USB. DO NOT CHANGE THE NAME FROM LOGO

Turn OFF your unit, insert the USB into the USB port located on the back of the unit.

Turn the unit back ON/

Please wait until the unit has completed the boot up and warm up process

 

520 and 720:

Press  SETUP and go to option (4) RESULT SHEETS

Change the logo setting to BMP LOGO and press exit a couple times and save your setting.

 

230:

Press option (3) RESULTS and change the LOGO TYPE to BMP LOGO

 

270/570/770:

Press the ADMINISTRATOR MENU Button and go to option (14) and choose the Image option.

 

370:

1. Press the SETUP on the top left Corner

2. Press the Result Sheet Options Menu

3. Go to Result Sheet Custom Logo, select custom logo and press the SAVE button.

Can I change or disable the sound/music?

230:

Press SETUP on the keypad

Go to option 4. OTHERS

Scroll down to change the volume and/or sound tone of the InBody unit.

520 and 720:

Press SETUP on the keypad

Go to option 6. OTHERS

Scroll down for the options to change the volume and/or sound type

 

370:

Press SETUP on the top left corner of the main screen

 

Press SETTINGS and go to Language Sound Type/ Volume

 

570/770:

Press the ADMINISTRATOR MENU button on the bottom left corner

Press option (01) Date/Time/Units/Password/Volume

How do I change the default gender?

520 and 720:

Press SETUP and go to option (6) OTHERS. Scroll down on the bottom and change the DEFAULT GENDER

 

270 570 770:

Under the ADMINISTRATOR MENU go to option (05.) Bypass Age/Gender. Select default gender

How do I clear/reset my database?

230:

Press SETUP on your Keypad

Go to options 4. Results

Use the keypad to highlight option (4) Data Clear

Once highlighted, press ENTER on your keypad to clear the database history

 

520 and 720:

Press SETUP on your Keypad

Go to option 6. Others

Scroll down using the arrow keys and highlight INITIALIZE HISTORY

Press ENTER to clear the database permanently

 

370:

Press the DATABASE button on the top left corner

Press SELECT ALL and then DELETE

 

570/770:

Press the ADMINISTRATOR MENU button on the bottom left corner

Press Option (6) to view the database

Press the top circle on the right hand side to highlight all the data

Once highlighted, press the DELETE button on the bottom right to clear your database.

How do I export/copy the database on to an Excel file?

230:

Insert your InBody USB thumb drive into the USB port and press ENTER on your keypad

A sub menu will appear, please press ENTER for yes

 

520: 

Insert your InBody USB thumb drive into an open USB port

Press DATABASE on your keypad

Once in the Database, press SETUP on your keypad

Choose (1) to back up the database file

Choose(3) to copy the database file onto an Excel spreadsheet

 

720: 

Insert your InBody USB thumb drive into an open USB port

Press DATABASE on your Keypad to access the database

Below the database screen, you will see options to save, backup, or load the database

A sub menu will appear: (1) Backup (2) Restore (3) Copy.  Please select option (3) copy to view your database on an Excel spreadsheet.

 

370:

Insert your InBody USB thumb drive into the USB Port

Press DATABASE on the top left corner

To copy the database as an Excel spreadsheet, press SELECT ALL then COPY

To backup the database, please press BACKUP

 

270/570/770

Insert your InBody USB thumb drive into the USB port on the back of the InBody

Press the ADMINISTRATOR MENU button on the bottom left corner

Go to Option (7) Export Data as Excel

You can  choose export all of the data or select individual tests

Once selected, press start

**Note: It may take several hours when exporting large amounts of tests** 

Can I take my InBody outside of the US?

Please refer to your contract. If you take your InBody outside of US territory, InBody is no longer obligated to service your unit and the warranty will become void.

How do I load or restore the database onto the InBody?

In order to restore or load the database, you will need the database or backup file.

*NOTE* 


Restoring the database file on the 520/720 will replace the current data on your unit

For the 370/570/770, unless you have the “combine” option, it will also replace the current data on your unit. Please contact our support department if you have any inquires prior to restoring the database onto your machine.

Transfer the backup file into the InBody USB thumb drive.

The file will have a .bin or .bak extention.  Once you have loaded the file onto your InBody USB thumb drive, please follow the instructions accordingly to your InBody unit.

520:
1. Insert USB into any open USB port on the back if the InBody

2. Press DATABASE on your keypad

3. Press SETUP and chose option 2 RESTORE

370:

1. Insert USB into any open USB port on the back if the InBody

2. Press DATABASE on the touch screen

3. Press RESTORE

720:

1. Insert USB into any open USB port on the back if the InBody

2. Press DATABASE on the keypad

3. Press DATABASE again to access the backup/restore menu

4. Press 2 RESTORE

570/770:

1. Insert USB into any open USB port on the back if the InBody

2. Press the ADMINISTRATOR MENU on the screen

3. Go to option 08. BACKUP/RESTORE ( some units may have a COMBINE option)

4. Choose RESTORE DATA and then START

5. You can also choose DATA COMBINE to add on to your current database

The result sheet is printing an error message about corrupted data

This issue is due to the printer setting and not the InBody.  If you are using a Samsung printer, please go to SETUP, then printer type, and change the setting from SPL-2009 to SPL2009-600.  If the setting is already set to SPL 2009-600, please change it to SPL-2009.